Component: nanoElute system (SN1837934524729)
Framework: nanoElute V1.1.0.63 (356, 2109) 
Simulation: False 
---------------------------------------------------------------
Component: Gradient pump
Driver: 0.9.9.25 
Firmware: 183 
Hardware: 7 
Serial number: 210039 
Flow offset A: 0 µL/min
Flow factor A: 1 
Flow offset B: -0.0073 µL/min
Flow factor B: 2.393358 
Pressure offset A: 125.0879 
Pressure factor A: 0.01972 
Pressure offset B: 128.3378 
Pressure factor B: 0.01972 
Valve shifts A: 23133 
Rotor shifts A: 12024 
Syringe mileage A: 223851 mm
Seal mileage A: 128619 mm
Valve shifts B: 21715 
Rotor shifts B: 11188 
Syringe mileage B: 254041 mm
Seal mileage B: 152308 mm
---------------------------------------------------------------
Component: Autosampler
Driver: 2.3.2 (2.3.16075.643) 
Firmware: 2.2.9 (2.2.17045.1553) 
---------------------------------------------------------------
Component: Journal
--- Flow restriction test: 12/21/22 15:44:06 
FR: Route 1 [µL/min]: 0.631796666666667 
FR: Route 2 [µL/min]: 1.76653709919453 
FR: Route 3 with loop [µL/min]: 4.32073709623715 
--- Leak test: 12/21/22 15:56:38 
LT: Pump A head [µL/min]: 0.0347669586997768 
LT: Pump B head [µL/min]: 0.614076644021566 
LT: Route 1 tubing [µL/min]: -0.000343333333333333 
LT: Route 2 tubing [µL/min]: 0.000390656248728433 
LT: Route 3 with loop [µL/min]: 0.000478671550750732 
--- Preparation: 01/23/23 10:21:40 
Preparation A [µL]: 7.6074713304228 
Preparation B [µL]: 15.5069013381192 
---------------------------------------------------------------