Component: nanoElute system (SN1837934494867)
Framework: nanoElute V1.1.0.63 (356, 2109) 
Simulation: False 
---------------------------------------------------------------
Component: Gradient pump
Driver: 0.9.9.25 
Firmware: 182 
Hardware: 7 
Serial number: 219050 
Flow offset A: 0.002 µL/min
Flow factor A: 1 
Flow offset B: 0.0003 µL/min
Flow factor B: 2.3 
Pressure offset A: 124.0072 
Pressure factor A: 0.01972 
Pressure offset B: 130.7988 
Pressure factor B: 0.01972 
Valve shifts A: 44824 
Rotor shifts A: 15520 
Syringe mileage A: 519312 mm
Seal mileage A: 519312 mm
Valve shifts B: 40050 
Rotor shifts B: 40050 
Syringe mileage B: 540020 mm
Seal mileage B: 540020 mm
---------------------------------------------------------------
Component: Autosampler
Driver: 2.3.2 (2.3.16075.643) 
Firmware: 2.2.9 (2.2.17045.1553) 
---------------------------------------------------------------
Component: Journal
--- Leak test: 06/03/24 14:49:50 
LT: Pump A head [µL/min]: 0.115610609652094 
LT: Pump B head [µL/min]: 1.49456034506773 
LT: Route 1 tubing [µL/min]: -0.000636666666666666 
LT: Route 2 tubing [µL/min]: 0.0077366665124893 
LT: Route 3 with loop [µL/min]: 0.0232759995174408 
--- Preparation: 06/03/24 16:11:19 
Preparation A [µL]: 7.28619112171569 
Preparation B [µL]: 15.987146135098 
--- Flow restriction test: 06/05/24 08:56:53 
FR: Route 1 [µL/min]: 1.00293 
FR: Route 2 [µL/min]: 2.21418028743506 
FR: Route 3 with loop [µL/min]: 6.47838293072255 
---------------------------------------------------------------